Human Acid Fibroblast Growth Factor (FGF-acid) ELISA Kit Instructions for Use

Human acidic fibroblast growth factor (FGF-acid) ELISA kit
  ( used in serum, plasma, cell culture supernatants and other biological fluids )
principle
This experiment used double antibody sandwich ABC-ELISA. The anti-human FGF-acid monoclonal antibody is coated on the microplate, the FGF-acid in the standard and the sample is combined with the monoclonal antibody, and biotinylated anti-human FGF-acid is added to form an immune complex attached to the plate. Horseradish peroxidase-labeled Streptavidin is combined with biotin, the substrate working solution is blue, and finally the stop solution sulfuric acid is added. The OD value is measured at 450 nm. The FGF-acid concentration is directly proportional to the OD value, which can be drawn by standard. The curve determines the concentration of FGF-acid in the specimen.
Kit composition ( 2-8 ° C preservation)
Coated Wells
96 holes
Enzyme Conjugate
12ml
10× specimen dilution (Sample Buffer)
12ml
20×Wash Buffer
50ml
Standards: 4ng/bottle
2 bottles
Substrate working fluid (TMB Solution)
12ml
Primary antibody working solution (Biotinylated Antibody)
12ml
Stop Solution
12ml
Prepare reagents and collect blood samples
1. Collection of specimens: serum, plasma (EDTA, citrate, heparin anticoagulation), cell culture supernatant, tissue homogenate, etc., as early as possible, stored at 2-8 ° C for 48 hours; longer time must be frozen (-20 °C or -70 °C) Save to avoid repeated freezing and thawing.
2. Standard solution preparation: Add 1ml of distilled water before use and mix well to form a solution of 4000pg/ml. Set 8 tubes of standard tubes, and add 300 ul of standard dilution solution to each tube. Add 300 ul of the standard solution of 4000 pg/ml to the first tube, mix it, and then suck 300 ul with the sampler and transfer to the second tube. Repeat the dilution in this way, and remove 300 ul from the seventh tube and discard it. The eighth tube is a blank control.
3. The 10× specimen dilution was diluted 1:10 with distilled water (example: 1 ml concentrated dilution + 9 ml distilled water).
4. Washing solution: diluted 1:20 with distilled water (example: 1 ml concentrated washing solution added to 19 ml of distilled water)
Test procedure
1. Loading: Add 100 ul of standard or sample to be tested in each well. Mix the reaction plate thoroughly and let it stand at 37 °C for 120 minutes.
2. Wash the plate: Wash the plate thoroughly with washing solution 4-6 times, and dry it on the filter paper.
3. Add 100 ul of the first antibody working solution to each well. The reaction plate was thoroughly mixed and placed at 37 ° C for 60 minutes.
4. Wash the board: the same as before.
5. Add 100 ul of enzyme-labeled antibody working solution per well. The reaction plate was placed at 37 ° C for 30 minutes.
6. Wash the board: same as before.
7. Add 100 ul of substrate working solution per well, set 37 The reaction was carried out in the dark at °C for 15 minutes.
8. Add 100 ul of stop solution to each well and mix.
9. Measure the absorbance at 450 nm using a microplate reader within 30 minutes.
Result calculation and judgment
1. All OD values ​​should be subtracted from the blank value before calculation.
2. Take the standard products 2000, 1000, 500, 250, 125, 62.5, 31.2, 0 pg/ml as the abscissa and OD as the ordinate. Draw on the coordinate paper and draw the standard curve.
3. Find the corresponding FGF-acid content on the graph based on the OD value of the sample.
Kit performance
1. Sensitivity: The minimum FGF-acid detection concentration is less than 16pg/ml.
2. Specificity: Recombinant or natural human FGF-acid can be detected simultaneously. Does not cross-react with other human cytokines.
3. Repeatability: The coefficient of variation in both the plate and the plate is less than 10%.
Precautions
1. It is recommended to make double holes for the above standard holes and samples to be tested. The standard curve should be made at the same time for each measurement.
2. The washing process is critical. Insufficient washing will result in an accuracy error and an erroneous rise in the OD value.
3. After the slats are opened, the remaining slats should be sealed again to keep the slats dry .
4. This kit should be stored in a 4 o C refrigerator.
5. This kit is for scientific research only and cannot be used for clinical diagnosis!

Thermometer

Thermal Imaging Thermometers are compact, have a focus-free lens and simple navigation for easy one-handed operation, tough enough for rugged field use. Temperature range is -4°-482°F (-20°-250°C) ±2% or 2°C and batteries last about four hours with quick charging available from the included AC charger. Models are rated to withstand shock values to 25g. Thermal image enhancement adds key details from the visible spectrum camera to the entire infrared image on the 3in color LCD screen, displaying accurate thermal gradients and visible details.

All models have 45° x 34° field of view and 320x240 pixel resolution for the image. Enhanced MA-776 and MA-778 models feature better infrared resolution, selectable measurement modes and picture-in-picture display, which superimpose a thermal image over a visible spectrum photo. Up to 500 sets of infrared, visual and MSX images can be stored as .jpeg files in internal memory for later analysis and reporting. The unmatched manufacturer`s warranty offers 10 year coverage on the detector, 5 year on the battery and 2 years parts and labor.

Clinical Thermometer,Thermometer,Nfrared Body Thermometer,Lab Classic Clinical Thermometer

Suzhou Xuanweicheng Biotechnology Co., Ltd , https://www.xwc-medical.com