Medical and Vet/Veterinary Volumetric Infusion Pump

Model NO.: MVP N
Purge: Yes
Bolus: Yes
Drug Library: No
Record: No
Kvo Rate: 0.1-5 Ml/H (in 0.1 Ml/H Increments)
Battery Life: Approx 5 Hours at 30ml/H Flow Rate
Weight: 2.5kgs
Trademark: winport
Transport Package: Carton
Specification: CE, ISO13485
Origin: China
HS Code: 90183900
Real-time infused volume / bolus rate / bolus volume / KVO rate, automatic power switching, mute key, purge, bolus, system memory, key locker, change flow rate without stopping the pump

Features:

• Compact in design, light in weight and small in size.
• Compatible with universal IV set.
• Low motor driving noise.
• Ultrasonic bubble sensor.
• Easy to set VTBI (volume to be infused) by [INCR] or [DECR] key on the front panel.
• Accurate setting of flow rate for patients.
• Flow rate accuracy with the equipped peristaltic finger system.
• The infused volume can be cleared by pressing [CLEAR] key without switching off the power.
• Audio-visual alarms for added safety.
• The reminder alarm sounds repeatedly if no action is taken within 2 minutes after the alarm was switched off.
• The flow rate can be set in 0.1ml/h increments.
• After delivering the VTBI, the pump continues running with keep vein open (KVO rate) mode.
• When door is open, the tube is automatically locked by tube clamp.
• The rechargeable built-in battery allows the pump to be transported with the patient without ceasing normal pump
operation.

Specifications

Model MVP N
Pumping Mechanism Curvilinear peristaltic
IV Set Compatible with IV sets of any standard
Flow Rate 0.1-1500 ml/h (in 0.1 ml/h increments)
Purge, Bolus 100-1500 ml/h (in 1 ml/h increments)
Purge when pump stops, bolus when pump starts
Bolus volume 1-20 ml (in 1 ml increments)
Accuracy ±3%
VTBI 1-9999 ml
Infusion Mode ml/h, drop/min, time-based
KVO Rate 0.1-5 ml/h (in 0.1 ml/h increments)
Alarms Occlusion, air-in-line, door open, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, reminder alarm
Additional Features Real-time infused volume / bolus rate / bolus volume / KVO rate,
automatic power switching, mute key, purge, bolus, system memory,
key locker, change flow rate without stopping the pump
Occlusion Sensitivity High, medium, low
Air-in-line Detection Ultrasonic detector
Wireless management Optional
Power Supply, AC 110/230 V (optional), 50-60 Hz, 20 VA
Battery 9.6±1.6 V, rechargeable
Battery Life 5 hours at 30 ml/h
Working Temperature 10-40ºC
Relative Humidity 30-75%
Atmospheric Pressure 700-1060 hpa
Size 174*126*215 mm
Weight 2.5 kg
Safety Classification Class I, type CF

Medical and Vet/Veterinary Volumetric Infusion Pump
1) LCD display
Medical and Vet/Veterinary Volumetric Infusion Pump

Medical and Vet/Veterinary Volumetric Infusion Pump

2) Keys
Medical and Vet/Veterinary Volumetric Infusion Pump
3) Flow LED: Flash when infusion.
4) DIP switch: IV set selection.
5) Bubble sensor: Detects air bubble in the IV tubing.
6) Tension plate: Press on the tubing to generate pressure.
7) Fingers: Peristaltic press on the tubing to drive down the solution.
8) Door lever: Pull up this lever and close the door. Just push the door to lock it.
9) Tubing guider: Guide the IV tubing.
10) Pressure sensor: Detects the occlusion of the IV tubing.
11) Tubing clamp: Automatically clamps the IV tubing when the door is open.
12) Pump door.

RF System

Slimming Machine,Beauty Machine,Skin Tightening

Lasy Science & Technology Co., Ltd. , http://www.ywelightmachine.com